Calibration / adjustment required
The balance needs to perform a calibration/adjustment (isoCAL). When the set time or a temperature change is exceeded, the balance requests an adjustment to maintain accuracy.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Empty the weighing pan.
- Start the calibration/adjustment by pressing the isoCAL / CAL button on the display.
- If an external weight is requested, use the calibration weight specified on the display.
Possible causes
- The isoCAL time interval has expired
- The ambient temperature changed
- The balance was moved
Do you need service?
The message clears once the user-initiated calibration/adjustment is done. No service is required.
Models showing this code: Secura
Source: Sartorius official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.