Code on display Leveling Sartorius Secura

Leveling process

This is a status message indicating that the balance needs to be leveled. On the Secura with its electronic level indicator, the sensors detect tilt, and even if the message is confirmed, it repeats every 2 minutes until the balance is leveled.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Start the leveling function by pressing the LEVEL button on the display.
  2. Turn the front feet in the direction of the arrow on the display until the animation turns green.
  3. Calibrate after leveling.

Possible causes

  • The balance is tilted
  • Change of location
  • The feet adjustment was disturbed

Do you need service?

Leveling is done by the user with the front feet. No service is required.

Models showing this code: Secura

Source: Sartorius official user manual