Initial or calibration mass out of range
The initial mass or the mass measured during calibration is outside the permitted range. The balance does not complete startup or calibration in this case.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Empty the weighing pan and check that it is installed correctly.
- Use the nominal weight requested by the balance for calibration.
- Restart the balance.
Possible causes
- A load was on the weighing pan at switch-on
- Wrong calibration weight
- The weighing pan is not installed
Do you need service?
If Er1 persists with an empty and correctly installed pan, technical service is needed for the load cell/calibration.
Models showing this code: AS
Source: RADWAG official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.