Code on display Er1 Hi RADWAG AS

Initial or calibration mass out of range

The initial mass or the mass measured during calibration is outside the permitted range. The balance does not complete startup or calibration in this case.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the weighing pan and check that it is installed correctly.
  2. Use the nominal weight requested by the balance for calibration.
  3. Restart the balance.

Possible causes

  • A load was on the weighing pan at switch-on
  • Wrong calibration weight
  • The weighing pan is not installed

Do you need service?

If Er1 persists with an empty and correctly installed pan, technical service is needed for the load cell/calibration.

Models showing this code: AS

Source: RADWAG official user manual