Code on display -LH- RADWAG AS R2

Initial mass error

At startup the balance compares the weight of the pan and its holder with the stored initial mass. If the difference exceeds the permitted range, LH is displayed. Typical causes are a load on the pan or a pan that is not installed.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Turn the balance off and empty the pan.
  2. Check that the weighing pan is installed correctly.
  3. Turn the balance back on.

Possible causes

  • Load on the weighing pan at startup
  • The weighing pan is not installed or is seated incorrectly

Do you need service?

If LH persists with an empty, correctly installed pan, the initial mass setting or the measuring system must be checked by technical service.

Source: RADWAG official user manual · Open manufacturer document