Outside zeroing range
According to the manual, zeroing can only be done within ±2% of maximum capacity. When the display is outside this limit, the balance refuses to zero. This limit prevents an actual load on the pan from being zeroed by mistake.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Empty the weighing pan completely.
- Check that the weighing pan is seated correctly.
- Press the zero key again once the indication is stable.
- If you are using a container, tare instead of zeroing.
Possible causes
- Sample or container on the pan
- Pan installed incorrectly
- An object touching the pan
Do you need service?
If Err2 continues with the pan empty and correctly installed, the balance's zero point must be checked by technical service.
Source: RADWAG official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.