Overload
The load on the platform is above the upper limit of the scale. The scale does not show this load.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Remove the load from the platform.
- Split the load into portions within capacity.
- Check that nothing else touches the platform.
- A load above capacity can permanently damage the load cell.
Possible causes
- A load heavier than the capacity
- An object leaning on the platform
Do you need service?
If Err 8.3 appears with an empty platform, the load cell must be checked by a technician.
Source: OHAUS official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.


