Code on display Err 8.1 OHAUS Valor 7000

Power-on zero limit exceeded

The scale sees more weight than expected on the platform at start-up and cannot set its zero point.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove the load from the platform.
  2. Check that nothing touches the platform.
  3. Switch the balance off and on again.

Possible causes

  • Load on the platform at power-up
  • An object touching the platform

Do you need service?

If Err 8.1 remains with an empty platform, the load cell must be checked by a technician.

Source: OHAUS official user manual · Open manufacturer document