Code on display Err 8.2 OHAUS Ranger 3000 / R31

Below power-on zero limit

If the pan is not installed or is seated incorrectly, the power-up signal is lower than expected. If it also appears with the pan installed correctly, the calibration should be checked.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the pan is in place and seated correctly.
  2. Switch the balance off and on again.
  3. If the error continues, calibrate the scale.

Possible causes

  • The weighing pan is not installed
  • The weighing pan is seated incorrectly
  • Calibration drift

Do you need service?

If it continues with the pan correct and after calibration, a service check is required.

Source: OHAUS official user manual · Open manufacturer document