Code on display Err 8.1 OHAUS Ranger 3000 / R31

Power-on zero limit exceeded

The scale detects a load on the pan at power-up and cannot take the initial zero. If it also appears with an empty pan, the calibration may have drifted.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the weighing pan.
  2. Check that nothing is touching the weighing pan.
  3. Switch the balance off and on again.
  4. If the error continues, calibrate the scale.

Possible causes

  • There was a load on the weighing pan at switch-on
  • An object is touching the weighing pan
  • Calibration drift

Do you need service?

If it continues with an empty pan and after calibration, contact technical service to check the load cell.