Code on display tiMout OHAUS PR Series Precision

Operation timeout

According to the manual, the cause is an unstable environment. Time runs out while the balance waits for a stable reading.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Move the balance to a solid, vibration-free bench.
  2. Avoid drafts from air conditioning, doors and windows.
  3. Make sure the sample sits still on the pan.

Possible causes

  • Vibration or air drafts
  • Unstable or shaky table
  • Sample moving on the pan

Do you need service?

If tiMout remains in a stable environment, the stability of the weighing cell must be checked by a technician.

Models showing this code: PR4202, PR6201