Calibration failed
The balance could not get a valid value during calibration. The manual lists three causes: unstable environment, wrong weight and system error.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Move the balance to a suitable location and repeat calibration.
- Use weights that match the calibration points in the specification table.
- Restart calibration with an empty pan.
Possible causes
- Unstable environment (vibration, drafts)
- Wrong calibration weight
- System Error
Do you need service?
If CAL.FAIL remains with the correct weight in a stable place, there is a system error and the balance must be checked by a technician.
Models showing this code: PR4202, PR6201
Source: OHAUS official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.