Code on display InVALid OHAUS PR Series Precision

Invalid reference value

It appears when the average piece weight is below 0.1d in parts counting, or the reference weight is below 100d in percent weighing. The balance cannot give a reliable result with this reference.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Recalculate the average piece weight (APW) in parts counting.
  2. If the error remains, use a different sample.
  3. Place more sample on the pan in percent weighing.

Possible causes

  • Average piece weight too small
  • Reference weight too low in percent weighing

Do you need service?

This message relates to the sample weight and does not indicate a fault in the balance.

Models showing this code: PR4202, PR6201

Source: OHAUS official user manual · Open manufacturer document