Code on display --NO-- OHAUS PR Series Analytical

Action not possible

The balance refuses the action if the value is outside the zero range or the tare value is negative. The zero range depends on the region setting.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Press Zero instead of Tare on a negative value.
  2. Clear the pan and zero again.
  3. Check that you are within the zero range for your region setting.

Possible causes

  • Value outside the zero range
  • Tare pressed on a negative value

Do you need service?

This is an operating notice and clears when the correct key is used.

Models showing this code: PR224, PR224/E, PR223

Source: OHAUS official user manual · Open manufacturer document