Code on display CAL.FAIL OHAUS PR Series Analytical

Calibration failed

The balance could not get a valid value during calibration. The manual lists three causes: unstable environment, wrong weight and system error.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Move the balance to a suitable location and repeat calibration.
  2. Use weights that match the calibration points in the specification table.
  3. Restart calibration with an empty pan.

Possible causes

  • Unstable environment (vibration, drafts)
  • Wrong calibration weight
  • System Error

Do you need service?

If CAL.FAIL remains with the correct weight in a stable place, there is a system error and the balance must be checked by a technician.

Models showing this code: PR224, PR224/E, PR223