Below power-on zero range
The balance does not see the expected pan weight at start-up. This mostly happens when the pan is not installed.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Install the pan before power-on.
- Use the original OHAUS pan for your model.
- Switch the balance off and on again.
Possible causes
- The weighing pan is not installed
- Wrong or non-original weighing pan
Do you need service?
If Err 8.2 remains with the correct pan installed, the weighing cell must be checked by a technician.
Models showing this code: PR224, PR224/E, PR223
Source: OHAUS official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.