Code on display Err 53 OHAUS Navigator NVT

Invalid checksum data

Settings or calibration data in memory may be corrupted. Correct weighing cannot be expected with this data.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Turn the balance off, unplug it and switch it on again after a short wait.
  2. Check whether the error returns at every start-up.
  • Do not rely on weighing results while Err 53 is shown.

Possible causes

  • Corrupted memory data
  • Electronics board failure

Do you need service?

The stored data must be reloaded or the memory board repaired by a technician.

Models showing this code: NVT6201, NVT10201

Source: OHAUS official user manual · Open manufacturer document