Temperature above setting
According to the service manual, this error appears when the temperature is 20 °C above the target. Do not start a new test until the analyzer has cooled down.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Open the cover and let the analyzer cool down.
- Start the test again once it has cooled down.
- The heater and sample chamber may be very hot. Take care when opening the cover.
Possible causes
- Test started at a lower temperature right after a high-temperature test
- Analyzer not cooled down enough
Do you need service?
If Err 8.6 keeps returning after cooling, the temperature control must be checked by a technician.
Models showing this code: MB23, MB25, MB27
Source: OHAUS official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.