Code on display Err 53 OHAUS MB23 / MB25 / MB27

EEPROM checksum error

Stored settings may be corrupted. The service manual recommends switching the unit off and on as a first step.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Turn the device off.
  2. Wait a few seconds and turn it back on.
  • Do not trust measurement results while Err 53 is shown.

Possible causes

  • Corrupted memory data
  • Faulty cable between load cell and main board
  • Main board fault

Do you need service?

If the error remains, the cable connection and main board must be checked by a technician.

Models showing this code: MB23, MB25, MB27

Source: OHAUS official user manual · Open manufacturer document