Code on display UNLOAD KERN & Sohn PLS

Weighing range lower limit error

The weighing range was not reached. The measured value is below the lower limit. The manual recommends checking the weighing pan position.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Check that the weighing pan and pan support are installed.
  2. Seat the weighing pan correctly.
  3. Turn the device off and on again after 30 seconds.

Possible causes

  • The weighing pan is not installed
  • Weighing pan or pan support seated incorrectly

Do you need service?

The warning clears when the weighing pan is placed correctly.

Models showing this code: PLS

Source: KERN & Sohn official user manual · Open manufacturer document