Code on display ERR01 KERN & Sohn PLS

Unstable weighing or zeroing error

The balance cannot zero because the weighing value has not stabilized. The manual recommends checking the environmental conditions.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Empty the weighing pan and make sure nothing touches it.
  2. Reduce air currents and vibration (close the glass doors).
  3. Repeat zeroing after the value stabilizes.

Possible causes

  • Air drafts
  • Table or floor vibration
  • An object is touching the weighing pan
  • Temperature change

Do you need service?

The error clears when the environment is stabilized. If it persists, a service check is recommended.

Models showing this code: PLS

Source: KERN & Sohn official user manual · Open manufacturer document