Code on display ERR03 KERN & Sohn PLJ

Wrong calibration weight

An error occurred during adjustment. The typical cause is using the wrong adjustment weight. The adjustment is not completed and the previous adjustment is kept.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Check the adjustment weight specified for your model in the manual.
  2. Repeat the adjustment with the correct weight.

Possible causes

  • Adjustment weight of the wrong value
  • Weight placed on the weighing pan too late

Do you need service?

The error clears when the correct adjustment weight is used.

Models showing this code: PLJ

Source: KERN & Sohn official user manual · Open manufacturer document