Code on display 1-Err KERN & Sohn PES

Adjustment weight too light

If the adjustment weight is less than 50 % of the weighing capacity, the balance rejects it. When calibrating the internal weight with an external one, the weight must be at least 95 % of the weighing range.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Use an adjustment weight as close as possible to the weighing capacity.
  2. Place the weight in the centre of the pan.
  3. Restart the adjustment.

Possible causes

  • Adjustment weight too light for the capacity
  • Wrong weight selected

Do you need service?

1-Err clears when adjustment is repeated with a suitable weight.

Source: KERN & Sohn official user manual · Open manufacturer document