Code on display A-Err KERN & Sohn PEJ

Internal adjustment ended faulty

At the end of the adjustment with the built-in weight, the balance did not get a valid result. The manual asks you to repeat the adjustment.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the weighing pan.
  2. Eliminate sources of vibration and drafts.
  3. Repeat the internal adjustment.

Possible causes

  • Vibration or draught during adjustment
  • Problem in the internal weight mechanism

Do you need service?

If A-Err remains after repeated adjustments, the internal adjustment mechanism must be checked by a technician.

Source: KERN & Sohn official user manual · Open manufacturer document