Code on display 4-Err KERN & Sohn PEJ

High deviation in ınternal adjustment

The balance compares the internal adjustment result with the expected value. If the difference is large, the adjustment is not considered reliable.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Leave the balance on for a while to warm up.
  2. Eliminate sources of vibration and drafts.
  3. Repeat the internal adjustment.

Possible causes

  • Vibration or air drafts
  • Balance not warmed up

Do you need service?

If 4-Err repeats in stable conditions, the internal adjustment mechanism must be checked by a technician.

Source: KERN & Sohn official user manual · Open manufacturer document