Code on display 2-Err KERN & Sohn PEJ

High deviation in adjustment test

The balance compares the value measured with the external weight to the expected value. If the difference exceeds 1 %, it rejects the result.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Use a correct adjustment weight close to the capacity.
  2. Check that the weight is clean and undamaged.
  3. Repeat the test.

Possible causes

  • Wrong or too light adjustment weight
  • Dirty or damaged weight
  • Unstable ambient conditions

Do you need service?

If 2-Err remains with the correct weight, the balance adjustment must be checked by a technician.

Source: KERN & Sohn official user manual · Open manufacturer document