Code on display ERR08 KERN & Sohn ALJ

Internal calibration error

An error occurred during internal adjustment (internal calibration). The balance could not complete adjustment with the internal weight.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the weighing pan.
  2. Reduce air drafts and vibration.
  3. Restart the internal adjustment.
  • Do not open the device.

Possible causes

  • Load left on the weighing pan during adjustment
  • Unstable environment
  • Internal adjustment mechanism failure

Do you need service?

If the internal adjustment fails again under suitable conditions, service is needed for the internal weight mechanism.

Models showing this code: ALJ

Source: KERN & Sohn official user manual · Open manufacturer document