Zero shift during adjustment
At the start of adjustment the balance reads a zero value with an empty pan. If this value is far from the expected one, adjustment is cancelled. The message also appears if nothing is done for 60 seconds during adjustment.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Press ON/OFF to return to weighing mode.
- Empty the pan and check that it is installed correctly.
- Restart the adjustment.
- Place the adjustment weight on the centre of the pan within 60 seconds.
Possible causes
- Objects on the pan when adjustment starts
- The weighing pan is not installed
- No operation for 60 seconds during adjustment
Do you need service?
If ERR C repeats with an empty, correctly fitted pan, the weighing cell must be checked by a technician.
Models showing this code: ACS 80-4, ACS 100-4, ACS 200-4, ACS 300-4
Source: KERN & Sohn official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.