Code on display ERR C KERN & Sohn ABJ

Zero shift during adjustment

At the start of adjustment the balance reads a zero value with an empty pan. If this value is far from the expected one, adjustment is cancelled. The message also appears if nothing is done for 60 seconds during adjustment.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Press ON/OFF to return to weighing mode.
  2. Empty the pan and check that it is installed correctly.
  3. Restart the adjustment and complete the steps without delay.

Possible causes

  • Objects on the pan when adjustment starts
  • The weighing pan is not installed
  • No operation for 60 seconds during adjustment

Do you need service?

If ERR C repeats with an empty, correctly fitted pan, the weighing cell must be checked by a technician.

Source: KERN & Sohn official user manual · Open manufacturer document