Code on display Err6 Jadever JWE

Incorrect calibration weight

The scale did not accept the load it read in the calibration step and did not complete calibration. The manual recommends trying again with the correct calibration weights and contacting service if that fails.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Use a weight of the correct class equal to the set calibration value.
  2. Check that the calibration value does not exceed the scale capacity.
  3. Empty the weighing pan and repeat the calibration from the start.
  • On legal-for-trade (approved) scales, calibration must be performed only by an authorized person.

Possible causes

  • Calibration weight of the wrong value
  • Calibration value entered larger than the capacity
  • Vibration or air currents during calibration

Do you need service?

If calibration is still not accepted after retrying with the correct weight, a service calibration is required.

Models showing this code: JWE-3K, JWE-6K, JWE-15K, JWE-30K

Source: Jadever official user manual