Incorrect calibration weight
The scale did not accept the load it read in the calibration step and did not complete calibration. The manual recommends trying again with the correct calibration weights and contacting service if that fails.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Use a weight of the correct class equal to the set calibration value.
- Check that the calibration value does not exceed the scale capacity.
- Empty the weighing pan and repeat the calibration from the start.
- On legal-for-trade (approved) scales, calibration must be performed only by an authorized person.
Possible causes
- Calibration weight of the wrong value
- Calibration value entered larger than the capacity
- Vibration or air currents during calibration
Do you need service?
If calibration is still not accepted after retrying with the correct weight, a service calibration is required.
Models showing this code: JWE-3K, JWE-6K, JWE-15K, JWE-30K
Source: Jadever official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.