Code on display Error1 Esit Smart-2

Weight upper limit exceeded

When the load exceeds the set capacity by 9 divisions (Max+9e), the Smart-2 does not show the weight and gives this code. Overload can permanently damage the load cells.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove the load from the platform immediately.
  2. Weigh the load by dividing it into portions that do not exceed the capacity.
  3. If the error continues on an empty platform, check the load cell cable connection.
  • Do not leave a load above capacity on the platform.

Possible causes

  • Load above capacity was applied
  • Capacity setting entered too low
  • Load cell or cable fault

Do you need service?

If Error1 also appears with the platform empty, there is a load cell or calibration problem and a service check is required.

Models showing this code: Smart-2

Source: Esit official user manual