Weight upper limit exceeded
When the load exceeds the set capacity by 9 divisions (Max+9e), the Smart-2 does not show the weight and gives this code. Overload can permanently damage the load cells.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Remove the load from the platform immediately.
- Weigh the load by dividing it into portions that do not exceed the capacity.
- If the error continues on an empty platform, check the load cell cable connection.
- Do not leave a load above capacity on the platform.
Possible causes
- Load above capacity was applied
- Capacity setting entered too low
- Load cell or cable fault
Do you need service?
If Error1 also appears with the platform empty, there is a load cell or calibration problem and a service check is required.
Models showing this code: Smart-2
Source: Esit official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.