A/D zero range error
The zero point measured at power-on is outside the zero range accepted by the A/D converter. In this case the balance cannot be zeroed and cannot weigh.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Empty the pan and seat it correctly.
- Check that nothing is touching the area around the pan.
- Switch the balance off and on again.
Possible causes
- Load left on the pan at power-on
- Pan installed incorrectly
- Load cell damaged by overload
Do you need service?
If the error persists with an empty and correctly installed pan, the load cell or calibration must be checked by service.
Models showing this code: KS-02
Source: Dikomsan official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.