Code on display O_Id Dikomsan KS-02

Overload

The load on the pan exceeds the maximum capacity of the scale. On precision balances, overload can permanently damage the load cell.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove the load from the pan immediately.
  2. Split the load so it stays below the capacity, then weigh it in parts.
  3. Check that the display returns to normal when the pan is empty.
  • Do not drop loads onto the pan.

Possible causes

  • Placing a load above capacity
  • Load dropped onto the pan with sudden impact
  • Load cell damage

Do you need service?

If O_Id still appears with an empty pan, the load cell may be damaged and a service check is needed.

Models showing this code: KS-02

Source: Dikomsan official user manual