Code on display E0.01 Detecto DL

Initial stability error

At power-on the scale waits for a stable reading to set the zero point. If the pan is moving or there is a product on it, startup cannot complete. The device warns you because all later weighings could be wrong.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Turn the device off and remove all products from the pan.
  2. Check that the pan is seated properly in its socket and does not wobble.
  3. Place the scale on a level surface free of vibration and away from air drafts.
  • Turn the device on again and do not touch the pan during power-on.

Possible causes

  • Product left on the pan
  • Pan not fully seated or wobbling
  • Air draft or vibration

Do you need service?

If the error repeats at power-on with an empty and stable pan, the load cell or pan mounting must be checked by service.

Models showing this code: DL

Source: Detecto official user manual