Code on display E0.02 Detecto DL

Initial zero range exceeded

The weight measured at power-on deviates from the zero point recorded at calibration by more than allowed, so the scale cannot zero itself. It usually indicates there is a product on the pan or the pan is not in place.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Turn the device off and remove the products from the pan.
  2. Check that the pan is correctly mounted on its support bracket.
  3. Make sure nothing is rubbing under or around the pan.
  4. Turn the device on again.
  • Do not subject the pan to impact or overload.

Possible causes

  • Product on the pan at power-on
  • The weighing pan is not installed or is seated incorrectly
  • Object caught under the pan
  • Shift or damage in the load cell

Do you need service?

If the error persists with the pan empty and correctly installed, the zero point may have shifted. Contact service for a calibration or load cell check.

Models showing this code: DL

Source: Detecto official user manual