Initial zero range exceeded
The weight measured at power-on deviates from the zero point recorded at calibration by more than allowed, so the scale cannot zero itself. It usually indicates there is a product on the pan or the pan is not in place.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Turn the device off and remove the products from the pan.
- Check that the pan is correctly mounted on its support bracket.
- Make sure nothing is rubbing under or around the pan.
- Turn the device on again.
- Do not subject the pan to impact or overload.
Possible causes
- Product on the pan at power-on
- The weighing pan is not installed or is seated incorrectly
- Object caught under the pan
- Shift or damage in the load cell
Do you need service?
If the error persists with the pan empty and correctly installed, the zero point may have shifted. Contact service for a calibration or load cell check.
Models showing this code: DL
Source: Detecto official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.