Code on display Error39 Citizen CT Series

Calibration weight and clock error

The code indicates that Error 3 (calibration weight not placed or out of tolerance) is occurring together with an RTC error. In this case the manual says to contact technical service.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Turn the balance off and on again. If the error persists, contact service.
  • Do not open the device.

Possible causes

  • Wrong or late calibration weight
  • RTC circuit or backup battery problem

Do you need service?

Because the RTC fault and calibration error occur together, technical service must check it.

Models showing this code: CT 253, CT 603, CT 1603

Source: Citizen official user manual