Code on display Error 3 Citizen CT Series

Calibration weight error

The balance waits for the calibration weight for 60 seconds. If no weight is placed, or if it differs from the expected value, the calibration is canceled.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Restart the calibration.
  2. When the balance asks for it, place the calibration weight of the correct value on the pan.
  3. Check that the weight matches the value requested on the display.

Possible causes

  • Weight not placed within 60 seconds
  • Calibration weight of the wrong value
  • Air drafts or vibration

Do you need service?

If calibration with the correct weight is repeatedly rejected, contact technical service to check the measuring system.

Models showing this code: CT 253, CT 603, CT 1603

Source: Citizen official user manual