Initial zero out of range
The weight on the pan did not fit the zero range when the scale started. It can also appear if the pan is not installed correctly or the environment is unsuitable.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Remove everything from the pan.
- Seat the pan correctly.
- Switch off and power on again with an empty pan.
Possible causes
- Load on the pan at power on
- Pan not seated correctly
- Air drafts or vibration
Do you need service?
If E2 remains with an empty pan, the load cell must be checked by a technician.
Models showing this code: XE-300, XE-600, XE-1500, XE-3000, XE-6000
Source: CAS official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.