Code on display ADM NOT IN CAS CL3000

Weighing module not responding

On the 7-segment display CODE 01 and 02 mean communication error, CODE 03 initialization error and CODE 04 setting value error. These point to an internal module problem.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Switch off, wait 30 seconds and switch on again.
  2. Note the CODE number shown on the display.
  • Do not open the scale housing.

Possible causes

  • A/D module communication error
  • Module failed to initialize
  • Corrupted module settings

Do you need service?

If ADM NOT IN returns after restart, the A/D weighing module must be checked by a technician.

Models showing this code: CL3000-B, CL3000-P

Source: CAS official user manual