Overload
The load on the platform exceeded the weighing capacity of the device. Because overload can permanently damage the load cell, the load must be reduced immediately.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Reduce or remove the load on the platform immediately.
- Weigh the load in parts so that it does not exceed capacity.
- Check whether the indicator returns to normal with the platform empty.
- Do not drop loads onto the platform.
Possible causes
- Placing a load above capacity
- Dropping the load with impact
- Load cell or device fault
Do you need service?
If Over also appears with the platform empty, the load cell may be damaged and a service check is needed.
Models showing this code: BX6, BX6W
Source: Baykon official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.