Code on display H AXIS AGN/C

Overload

The scale is overloaded, meaning the load on the weighing pan exceeds capacity. If it continues after the load is removed, there may be mechanical damage.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove the load from the weighing pan.
  2. Weigh the load again within the capacity limit.
  • Do not place a load above capacity on the weighing pan.

Possible causes

  • Placing a load above capacity
  • Mechanical damage to the scale

Do you need service?

If the H message also appears with an empty weighing pan, there may be mechanical damage. Technical service should check it.

Models showing this code: AGN50C, AGN100C, AGN200C

Source: AXIS official user manual