Code on display C-3 AXIS AGN/C

Startup self-test error

One of the self-tests run at startup gave a negative result. According to the manual, service is needed if the message stays on the display for more than 1 minute. It is not a fault if it appears briefly and disappears.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Wait about 1 minute for the self-test to finish.
  2. Make sure the weighing pan is empty and installed correctly, then turn the device off and on again.

Possible causes

  • A load was on the weighing pan at switch-on
  • Weighing pan not installed or installed incorrectly
  • Load cell or electronics fault

Do you need service?

If the C-x message stays for more than 1 minute, the self-test has found a fault and technical service is needed.

Models showing this code: AGN50C, AGN100C, AGN200C

Source: AXIS official user manual