Code on display ------ AXIS AGN/C

Tare cannot be completed

The tare (zeroing the weight reading) process cannot be completed. Tare cannot be taken because the scale cannot get a stable reading. The manual lists unstable positioning, floor vibration and air drafts as causes.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Move the scale to a stable table free of vibration.
  2. Keep it away from air drafts and close the draft shield doors.
  3. Press the tare key again.

Possible causes

  • Scale standing on an unstable surface
  • Floor vibration
  • Air drafts

Do you need service?

If tare still does not complete in a stable environment, the scale should be examined at technical service.

Models showing this code: AGN50C, AGN100C, AGN200C

Source: AXIS official user manual