Code on display 2-LocH / PrESS / 2Ero Avery Weigh-Tronix ZM305

Auto zero stability error

During automatic weighing, the indicator could not reach a stable zero weight within the set time window. The process does not continue.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the platform completely.
  2. Make sure the platform is stable and nothing is touching it.
  3. Reduce sources of vibration and zero with the ZERO key if needed.
  4. Restart the automatic weighing.

Possible causes

  • Material left on the platform
  • Vibration or air drafts
  • An object touching the platform

Do you need service?

If the error persists with an empty and stable platform, the zero stability must be checked by service.

Models showing this code: ZM305

Source: Avery Weigh-Tronix official user manual