Calibration failed
Span calibration could not be completed. The balance could not match the placed weight to the expected calibration value. Weighing results should not be trusted until calibration succeeds.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Empty the pan and restart calibration from the beginning.
- If the weight you are using differs from the value on the display, clear it with CE and enter the correct value.
- Place the weight in the center of the platform and complete calibration with the Tare key.
Possible causes
- Calibration weight used differs from the value shown on the display
- Another load on the pan during calibration
- Weight placed before a stable reading was obtained
Do you need service?
If repeated calibration with the correct weight still gives SPAn FAIL, the load cell or electronics should be checked by Adam dealer service.
Models showing this code: GFC
Source: Adam Equipment official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.