Code on display SPAn FAIL Adam Equipment GFC

Calibration failed

Span calibration could not be completed. The balance could not match the placed weight to the expected calibration value. Weighing results should not be trusted until calibration succeeds.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the pan and restart calibration from the beginning.
  2. If the weight you are using differs from the value on the display, clear it with CE and enter the correct value.
  3. Place the weight in the center of the platform and complete calibration with the Tare key.

Possible causes

  • Calibration weight used differs from the value shown on the display
  • Another load on the pan during calibration
  • Weight placed before a stable reading was obtained

Do you need service?

If repeated calibration with the correct weight still gives SPAn FAIL, the load cell or electronics should be checked by Adam dealer service.

Models showing this code: GFC

Source: Adam Equipment official user manual