A/D measurement error
At power-on, the count from the A/D converter is outside the normal range. This means the scale is not receiving a correct signal from the load cell.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Check that the weighing platform/pan is installed and seated correctly.
- Switch the device off and on again.
Possible causes
- Weighing platform not installed
- Load cell damage
- Electronics failure
Do you need service?
If Err 6 persists with the platform installed correctly, there is a load cell or electronics fault. Contact the service of the Adam dealer.
Models showing this code: GFC
Source: Adam Equipment official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.