Code on display -CAL E A&D HR-i

Calibration weight too light

The balance measured less than the set calibration value and did not complete calibration.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Press the CAL key to return to weighing mode.
  2. Compare your weight with the value on the display.
  3. Place the weight in the centre of the pan.
  4. Repeat the calibration with the correct weight.

Possible causes

  • Weight with the wrong nominal value
  • Weight not seated on the pan
  • Wrong calibration value setting

Do you need service?

If the error continues with the correct weight, the balance adjustment must be checked by a technician.

Source: A&D official user manual · Open manufacturer document