Code on display EC,E02 A&D HR-i

Balance not ready

This reply is sent when a command arrives while the balance is not in weighing mode or is still re-zeroing. The manual advises adjusting the command delay.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Return the balance to weighing mode.
  2. Increase the delay between commands in the software.
  3. Send the command again after re-zero finishes.

Possible causes

  • Balance not in weighing mode
  • Command sent during re-zero
  • Commands sent too often

Do you need service?

If the error continues with a longer delay, the communication settings should be checked by a technician.