Code on display CAL E A&D HR-i

Calibration weight too heavy

The balance measured more than the set calibration value and did not complete calibration.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Press the CAL key to return to weighing mode.
  2. Compare your weight with the value on the display.
  3. Make sure only the weight is on the pan.
  4. Repeat the calibration with the correct weight.

Possible causes

  • Weight with the wrong nominal value
  • Something else on the pan with the weight
  • Wrong calibration value setting

Do you need service?

If the error continues with the correct weight, the balance adjustment must be checked by a technician.

Source: A&D official user manual · Open manufacturer document