Code on display EC,E00 A&D HR-AZ

Communication protocol error

A protocol error occurred in the serial communication between the balance and the computer or printer, and the transmitted data could not be read correctly. Weighing is not affected, only data transfer is disrupted.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Check that the baud rate, parity, and format on the balance and the computer are the same.
  2. Check that the connection cable is the correct type and firmly connected.
  3. Retry communication after matching the settings.

Possible causes

  • Baud rate, parity, or data format mismatch
  • Faulty or loose connection cable

Do you need service?

The error clears once the communication settings on the balance and the computer are matched. The weighing function is not affected.

Models showing this code: HR-AZ

Source: A&D official user manual · Open manufacturer document