Code on display Error1 A&D HR-A

Weighing stability error

The balance cannot stabilize the reading because of an environmental factor and cannot complete the process (for example calibration or weighing by command). Vibration, air drafts, temperature changes, static electricity or magnetic fields cause the value to keep fluctuating.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Press the CAL key to return to weighing mode.
  2. Move the balance to a sturdy table free of vibration and away from air drafts.
  3. Close the draft shield doors if present, and keep statically charged or magnetic samples away from the weighing pan.
  4. Set the Response Adjustment to suit the environment and repeat the process.

Possible causes

  • Vibration on the table or floor
  • Air drafts from an air conditioner, door or window
  • Sudden temperature change
  • Static electricity or magnetic field in the sample

Do you need service?

If the balance still cannot give a stable reading after the environmental conditions are corrected, contact technical service to check the load cell and electronics.

Models showing this code: HR-A

Source: A&D official user manual · Open manufacturer document